Rigid CT / MR registration with SimpleITK

Reference scripts: rigid_ct_mr.py.

Multimodal rigid registration with SimpleITK

The problem in one line

Find the rigid transform T (3 rotations + 3 translations, Euler3DTransform) such that the moving image resampled through T lines up with the fixed image, when the two modalities do not share intensities: bone is bright in CT (+1000 HU) and dark in T1 MRI, CSF is dark in CT and bright in T2. Mean squares or correlation assume a linear intensity relation and fail; mutual information only assumes a statistical dependence between the two intensity distributions.

The stack, component by component

Component Setting Why
Transform Euler3DTransform initialised with CenteredTransformInitializer 6 DoF; the rotation centre at the fixed image centre (GEOMETRY) or the intensity centre of mass (MOMENTS) so rotations do not sweep the volume out of view
Metric MattesMutualInformation, 50 bins Parzen-window joint histogram; 32-64 bins is the sweet spot (fewer = smooth but blurry, more = noisy)
Sampling RANDOM, 1-15 % of voxels, fixed seed MI on all voxels is slow and over-fits noise; random sampling adds stochasticity that helps escape local minima
Optimizer RegularStepGradientDescent or GradientDescent (lr 1.0 then 0.2), SetOptimizerScalesFromPhysicalShift rotations (rad) and translations (mm) have different sensitivities: physical-shift scales equalise them
Multi-resolution shrink 4,2,1 with smoothing sigmas 2,1,0 mm capture-range at coarse levels, precision at fine; sigmas in physical units so anisotropic voxels are handled
Interpolator linear during optimisation, linear for the final image, nearest for labels, B-spline for the final image when quality matters
Convergence minimum value 1e-6 over a 10-iteration window, 100-500 iterations watch the metric curve: a plateau in the first 20 iterations means the capture range was missed

In pardesline every one of these is a parameter of simpleitk_registration.rigid:

import pardesline as pl

ds = pl.load("rire_brain")  # CT + MR-T1 of one brain, LPS mm
s = ds[0]
moved = s.ct.align(
    s.mr_t1,  # fixed = CT, moving = MR
    metric="mattes_mutual_information",
    number_of_histogram_bins=50,
    initializer="GEOMETRY",
    sampling_strategy="RANDOM",
    sampling_percentage=0.05,
    learning_rate=1.0,
    num_iterations=200,
    shrink_factors="4,2,1",
    smoothing_sigmas="2,1,0",
    smoothing_sigmas_in_physical_units=True,
    optimizer_scales_from_physical_shift=True,
    interpolator="LINEAR",
)
moved.transform  # rotation_deg, translation_mm, matrix_4x4, final metric

simpleitk_registration.advanced_rigid adds an exhaustive rotation search (exhaustive_steps_per_axis, phase 1) before gradient descent: use it when the initial misalignment exceeds ~30 deg or when the two scans were acquired in different patient positions. simpleitk_registration.affine adds scale and shear (12 DoF) for scanner calibration differences; bspline and demons are deformable (see the bottom of this page).

Frames: the part that breaks silently

  • DICOM and every ITK reader give LPS millimetres with origin, spacing and direction cosines; NIfTI headers are RAS and ITK converts them on read. pardesline's Volume.frame carries the affine; the registration operates in physical space, so two volumes with different spacing and orientation register without resampling first.
  • Never register voxel arrays (GetArrayFromImage) directly: you would lose the geometry and the transform would be meaningless in the scanner frame.
  • The output transform maps fixed physical points to moving physical points (ITK convention). To move an annotation from fixed to moving, apply T; to bring the moving image onto the fixed grid, Resample(moving, fixed, T). Label maps: nearest-neighbour interpolation, defaultPixelValue=0.
  • Composite registrations (rigid then B-spline): CompositeTransform, applied inner-to-outer; keep each stage's transform file (.tfm).

Diagnostics and validation

  1. Plot the metric per iteration (SimpleITK AddCommand(sitkIterationEvent)), check it decreases at each level and the level changes are visible.
  2. Checkerboard and colour overlay of fixed vs resampled moving (simpleitk_registration.difference_analysis gives a difference map).
  3. Target registration error on 5-10 anatomical landmarks placed independently in both volumes (not used by the algorithm): ||T(p_fixed) - p_moving|| in mm; sub-voxel on rigid brain CT/MR (RIRE reports ~1 mm).
  4. Dice of propagated masks when segmentations exist (skull, ventricles).
  5. Run twice with different random seeds / sampling: a stable result moves by less than the voxel size; otherwise increase sampling or bins.

Failure modes

Symptom Cause Fix
Metric flat, no motion capture range missed, initial misalignment too large MOMENTS initializer, advanced_rigid exhaustive search, coarser first level (shrink 8)
Converges to a wrong but plausible pose local minimum of MI (symmetric head, repeated vertebrae) multi-start, more bins, larger sampling, mask the field of view to the common anatomy
Drifts with iterations learning rate too high 0.2-0.5 with physical-shift scales; RegularStepGradientDescent with relaxation
Good at centre, bad at edges not rigid: scanner distortion, patient deformation affine then B-spline (grid_spacing 30-50 mm)
PET-CT off by a slice different acquisition times / breathing rigid on the torso mask only, then deformable
Works on brain, fails on abdomen organs move independently mask the organ of interest; deformable (demons for same-modality, B-spline MI for multimodal)

Going deformable

  • simpleitk_registration.bspline: B-spline FFD on a control-point grid (grid_spacing mm, multi-resolution scale_factors), LBFGSB, MI for multimodal. Validate with the Jacobian determinant (no folding) and TRE.
  • simpleitk_registration.demons: diffeomorphic demons on same-modality images (intensities must match), standard_deviations smooth the update field; fast, dense, not for CT-MR.
  • Always rigid (or affine) first; deformable models should correct millimetres, not centimetres.

Datasets to practise on

rire_brain (CT + MR-T1, the reference multimodal pair), popi_lung_4dct (same-modality breathing phases: rigid then B-spline/demons), ixi_brain_mri (T1 vs T2 vs PD of the same subject), chaos_abdomen (CT and MR of different patients: atlas-style registration), tcia_nsclc_radiomics_dicom (DICOM series as input), hf_openmind (T1 vs T2).

scripts/rigid_ct_mr.py registers the RIRE pair through pardesline, repeats the registration directly with SimpleITK to expose every setting, exports the checkerboard, and reports the metric trace and the transform.

Source: skills/multimodal-rigid-registration/SKILL.md, one of the twenty skills served by pardesline mcp; published on 2026-10-14.